-Applicable syringe size 2ml, 5ml, 10ml, 20ml, 30ml, 50/60ml
-Rate Range 2ml:0.01~60ml, 5ml:0.01~150ml, 10ml:0.01~300ml, 20ml:0.01~600ml, 30ml:0.01~900ml, 50/60ml:0.01~2200ml
-Rate Step 0.01ml/h
-BOLUS rate & volume 2ml:0.01~48ml/h 0.1~0.5ml 5ml:0.01~120ml/h 0.1~1ml 10ml:0.01~240ml/h 0.1~5ml 20ml:0.01~480ml/h 0.1~10ml 30ml:0.01~720ml/h 0.1~10ml 50/60ml:0.01~1500ml/h 0.1~10ml
-BOLUS rate step 0.01ml/h
-FF(Fast Forward) rate 2ml:0.01~60ml, 5ml:0.01~150ml, 10ml:0.01~300ml, 20ml:0.01~600ml, 30ml:0.01~900ml, 50/60ml:0.01~2200ml
-FF vol. 0 ~maximal volume of the syringe
-KVO rate 0.01~5.0ml/h (0.01ml step)
-KVO vol. 2ml:0~0.2ml, 5ml:0~0.2ml, 10ml:0~0.2ml, 20ml:0~0.5ml, 30ml:0~0.5ml, 50/60ml:0~1ml (0.01ml step)
-VTBI(volume to be infused) 0.01~9999.99ml, (0.01ml step)
-Cumulative inf. vol. 0.01~9999.99ml, (0.01ml step)
-Rate acc. Within ±2% Pressure units MPa, kPa, mmHg, inH2O, PSI, mbar
-Occlusion pressure Level 1-13 (10,20,30,40,50,60,70,80,90,100,110,120,130)kPa
-Alarms information infusion finished, near infusion end, VTBI finished, system stand-by completed, the infusion is blocked, syringe falls off, the syringe is improperly installed, forgotten operation, low battery, battery empty, abnormal running state, equipment failure, power failure, etc.
-Syringe max. value (under single fault condition) 1mL
-Power source, consumption, Fuse 100~240VAC, 50/60Hz, 25VA, T2A/250VAC Internal rechargeable Li battery 11.1V, capacity≥2600mAh, 8 hours backup time at 5ml/h External DC power supply, input voltage 15V, input current 1.2-1A.
-Equipment classification Class I, Type CF, Continuous operation
-IP class IP24
-Dim. & Net Weight 301mm(L)×183mm(W)×82mm(H) & 2.0kgs